Publication: Self-aligned czerny turner monochromator.
No Thumbnail Available
Date
2017-03
Authors
Wan Ab. Aziz, Nurul Ridzuan
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Automation have been a part of everyday necessity in the modern day industries. Assembly and process industries had started to replace the manual assembly processes with automated processes. These activities are also relevant in spectrometer industries. This is because, the manual way of performing optical alignment for a spectrometer is time consuming and taking a huge portion of total production time of one spectrometer. In this thesis, an automation system has been built that is able to perform the optical alignment automatically with high accuracy and precision, without a need of human supervision. There are two different algorithms has been programmed and tested for this system, first is X and Y Iteration Algorithm, and second is Look-Up Table Algorithm. Using the Look-Up Table algorithm, this system able to complete one of the alignment process within 14.26 seconds, compared to human which usually took more than 2 minutes to complete the same task. X and Y Iteration Algorithm in the other hand, needs longer time compared to Look-Up Table Algorithm, but produce higher accuracy results, with an error as low as 0.30 pixels, whereas Look-Up Table Algorithm has an error of 1.28 pixels. Overall, the results of the system performances have shown that it is capable of completing the optical alignment with high precision and accuracy.